Z-0990-2017 Class II Terminated

Recalled by Varian Medical Systems, Inc. — Palo Alto, CA

FDA device recall Z-0990-2017 was initiated by Varian Medical Systems, Inc. on Dec 2, 2016 and is designated Class II. Reason for recall: Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only… The recall status is terminated (terminated May 12, 2017). Affected quantity: 42 devices are affected.

Recall details

Product Type
Devices
Report Date
Jan 18, 2017
Initiation Date
Dec 2, 2016
Termination Date
May 12, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 devices are affected

Product description

Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring version 2.3 [2.3.12] Radiology: The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments.

Reason for recall

Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.

Distribution pattern

US Distribution to the states of : NJ, TN. OR and FL.

Code information

Model # H48 Serial numbers: H48L574 H48L575 H48L576 H48L577 H48L578 H48L579 H48L580 H48L581 H48L582 H48L583 H48L584 H48L585 H48L586 H48L587 H482484 H487302 H48A587 H48D957 H480113 H480118 H481884 H481885 H481886 H481916 H481917 H484198 H484199 H488949 H48A680 H48A871 H48A872 H48A873 H48A874 H48A875 H48A876 H48A877 H48A878 H48A879 H48I531 H481195 H481236 H481237.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0990-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Varian Medical Systems, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0990-2017. Report a data issue: hello@regdatalab.com.