Z-1903-2017 Class II Terminated

Recalled by Breg Inc — Carlsbad, CA

FDA device recall Z-1903-2017 was initiated by Breg Inc on Mar 30, 2017 and is designated Class II. Reason for recall: They may not have been sealed prior to sterilization. Products with unsealed pouches will be non-sterile. The recall status is terminated (terminated Jun 23, 2017). Affected quantity: 2,609 units total.

Recall details

Product Type
Devices
Report Date
May 10, 2017
Initiation Date
Mar 30, 2017
Termination Date
Jun 23, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,609 units total

Product description

Dressing Rect L, P/N 02428 Product Usage: These products are sterile dressings used as an insulation barrier between the Polar Care cold therapy pad and the skin. Sterile dressings are typically applied in a non-sterile environment after post-op and on occasion may be applied in a sterile environment. Breg has determined that a small percentage (approximately 0.22%) of these products may have been sent with unsealed pouches by the contract manufacturer. Customers will be contacted and instructed to inspect their inventory for any product with unsealed pouches and discard them.

Reason for recall

They may not have been sealed prior to sterilization. Products with unsealed pouches will be non-sterile.

Distribution pattern

US Nationwide Distribution

Code information

D163474, D163544

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1903-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.