Z-1905-2019 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1905-2019 was initiated by Siemens Medical Solutions USA, Inc on Jun 6, 2019 and is designated Class II. Reason for recall: Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to… The recall status is terminated (terminated Feb 9, 2021). Affected quantity: 8.

Recall details

Product Type
Devices
Report Date
Jul 10, 2019
Initiation Date
Jun 6, 2019
Termination Date
Feb 9, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8

Product description

Siemens Artis pheno-Interventional Fluoroscopic X-Ray System Model Number: 10849000

Reason for recall

Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel

Distribution pattern

Nationwide

Code information

Serial Numbers: 164021 164053 164055 164054 164088 164140 164113 164069

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1905-2019
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-1905-2019. Report a data issue: hello@regdatalab.com.