Z-1905-2019 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1905-2019 was initiated by Siemens Medical Solutions USA, Inc on June 6, 2019 and is designated Class II. Reason for recall: Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to th… The recall status is terminated (terminated February 9, 2021). Affected quantity: 8.

Recall Details

Product Type
Devices
Report Date
July 10, 2019
Initiation Date
June 6, 2019
Termination Date
February 9, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8

Product Description

Siemens Artis pheno-Interventional Fluoroscopic X-Ray System Model Number: 10849000

Reason for Recall

Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel

Distribution Pattern

Nationwide

Code Information

Serial Numbers: 164021 164053 164055 164054 164088 164140 164113 164069