Z-1905-2025 Class II

Recalled by VIRTUAL INCISION CORPORATION — Lincoln, NE

FDA device recall Z-1905-2025 was initiated by VIRTUAL INCISION CORPORATION on Apr 23, 2025 and is designated Class II. Reason for recall: The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between… Affected quantity: 3 units.

Recall details

Product Type
Devices
Report Date
Jun 11, 2025
Initiation Date
Apr 23, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product description

MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3

Reason for recall

The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.

Distribution pattern

US Distribution to States: CA, FL, NE

Code information

UDI-DI (01)00850038042028; Lot Numbers: 1020040211, 1020040212, 1020040213

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1905-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1905-2025. Report a data issue: hello@regdatalab.com.