Z-1906-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 21, 2023
- Initiation Date
- May 12, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,190
Product Description
Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body, REF: L-692020
Reason for Recall
A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosthetic system that replaces a missing lower limb, which can lead to falling and injury.
Distribution Pattern
US: SC, FL,NM, PA, VA, CA, TX, MT, UT, OH, NV, KS, MI, GA, IA, WA, TN, AZ, MN, NC, IN, MS, AL, NY, ID, DE, OK, MA, WI, OR, KY, MD, CO, WV, LA, IL, NJ, CT, MO, NE, WY, AR, SD, PR
Code Information
REF/UDI-DI: L-125000/05690967566142; L-621200/05690967596590; L-621000/05690967118334; L-621100/05690967155322; L-692020/05690967128548. Lots: HF210205, HF210519, HF211008, HF220303, MX210518, MX210816, MX210823, MX211022, MX220316.