Z-1911-2020 Class II Terminated

Recalled by ICU Medical, Inc. — San Clemente, CA

Recall Details

Product Type
Devices
Report Date
May 13, 2020
Initiation Date
October 4, 2019
Termination Date
December 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1175 units

Product Description

REF: AH7117 111" 10 Drop Admin Set w/2 Inj Sites, Rotating Luer, Sterile R, For Veterinary Use Only, (01) 1 0887709 051475

Reason for Recall

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

Distribution Pattern

US:AK,AL,AR,AZ,CA,CO,CT,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,SD,TN,TX,UT,VA,WA,WI,WY OUS: Canada *** "REF" numbers AH7117 and B30211 were not marketed or distributed within the US***

Code Information

Lot number 4173657 *Product not distributed within the US