Z-1912-2019 Class I Terminated

Recalled by Teleflex Medical — Morrisville, NC

Recall Details

Product Type
Devices
Report Date
July 24, 2019
Initiation Date
May 10, 2019
Termination Date
April 22, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 units

Product Description

Teleflex HUDSON RDI NEONATAL ConchaSmart Breathing Circuit with Dual Heated Limb and ConchaSmart Column: a) REF 870-07KIT b) REF 870-09KIT Product Usage: The Neonatal ConchaSmart breathing circuits are intended to deliver humidified breathing gases for administration to an infant/neonate patient. These gases may be delivered by nasal prongs using the CareFusion Infant Flow system and the single limb circuit, or through bypassing upper airways, for example through the use of an endotracheal tube with the dual limb configuration.

Reason for Recall

Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonatal ConchaSmart Breathing Circuit.

Distribution Pattern

US Nationwide Distribution in the states of AL, TX

Code Information

Lot/ Batch Number GTIN a) 74L1802045 14026704611154 b) 74L1802044 14026704611161