Z-1916-2025 Class II Ongoing
FDA device recall Z-1916-2025 was initiated by American Contract Systems, Inc. on Apr 21, 2025 and is designated Class II. Reason for recall: Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of… The recall status is ongoing. Affected quantity: 850 Kits.
Recall details
- Product Type
- Devices
- Report Date
- Jun 18, 2025
- Initiation Date
- Apr 21, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 850 Kits
Product description
LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE PACK, Model: NCTK31L All convenience kits contain Specialist Blend Cast Padding, Item: QTX30-322NS
Reason for recall
Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.
Distribution pattern
US Nationwide distribution in the state of CO.
Code information
Model/UDI-DI/Lot(Expiration): NCEX20I/00191072222398/9146304(2025-12-03), 9207685(2026-01-08); NCEX20J/00191072234582/9222096(2026-01-30), 9222096(2026-01-30); NCTK31K/00191072222480/9086225(2025-11-01), 9086225(2025-11-01); NCTK31L/00191072234872/9219596(2026-01-15), 9219596(2026-01-15)