Z-1922-2013 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 14, 2013
- Initiation Date
- July 24, 2013
- Termination Date
- June 13, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 37,310
Product Description
Nestle Nutrition, Compat, Enteral Feeding Tube, sterile, 8 French (REF 08310400), 10 French (REF 08310600), and 12 French (REF 08310500). Product Usage: The Compat Nasogastric Feeding Tube is indicated for the administration of nutrition, fluids and medications into the stomach r small bowel by the naso-enteric, route for patients who have an intact gastrointestinal tract but are physically unable to ingest enough nutrients orally to meet their current needs.
Reason for Recall
Nestle HealthCare Nutrition, Inc. is initiating a recall on some of Compat NG Tubes because they have been delivered in an insufficiently sealed pouch, thus compromising the sterility claimed on the label.
Distribution Pattern
US Nationwide Distribution - including the states of: AK, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA.
Code Information
REF 08310400 - Compat¿ Nasogastric Feeding Tube 8 French (8Fr) in the following lots: 01413G 02412K 08211Z 08712W 14212U 14611P 21512Y 35411H REF 08310500- Compat¿ Nasogastric Feeding Tube 12 French (12Fr) in the following lots: 01211AV 01513N 04111D, 04111P, 04111V 05812F 05912F 06313N 21512G 23412BV 34712V 35411U REF 08310600 - Compat¿ Nasogastric Feeding Tube 10 French (10Fr) in the following lots: 08111M 08712D 10211E 13712W 21512F 05312N