Z-1922-2013 Class II Terminated

Recalled by Nestle HealthCare Nutrition — Minnetonka, MN

Recall Details

Product Type
Devices
Report Date
August 14, 2013
Initiation Date
July 24, 2013
Termination Date
June 13, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
37,310

Product Description

Nestle Nutrition, Compat, Enteral Feeding Tube, sterile, 8 French (REF 08310400), 10 French (REF 08310600), and 12 French (REF 08310500). Product Usage: The Compat Nasogastric Feeding Tube is indicated for the administration of nutrition, fluids and medications into the stomach r small bowel by the naso-enteric, route for patients who have an intact gastrointestinal tract but are physically unable to ingest enough nutrients orally to meet their current needs.

Reason for Recall

Nestle HealthCare Nutrition, Inc. is initiating a recall on some of Compat NG Tubes because they have been delivered in an insufficiently sealed pouch, thus compromising the sterility claimed on the label.

Distribution Pattern

US Nationwide Distribution - including the states of: AK, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA.

Code Information

REF 08310400 - Compat¿ Nasogastric Feeding Tube 8 French (8Fr) in the following lots: 01413G 02412K 08211Z 08712W 14212U 14611P 21512Y 35411H REF 08310500- Compat¿ Nasogastric Feeding Tube 12 French (12Fr) in the following lots: 01211AV 01513N 04111D, 04111P, 04111V 05812F 05912F 06313N 21512G 23412BV 34712V 35411U REF 08310600 - Compat¿ Nasogastric Feeding Tube 10 French (10Fr) in the following lots: 08111M 08712D 10211E 13712W 21512F 05312N