Z-1926-2013 Class II Terminated

Recalled by SpineFrontier, Inc. — Beverly, MA

FDA device recall Z-1926-2013 was initiated by SpineFrontier, Inc. on Mar 14, 2013 and is designated Class II. Reason for recall: Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the… The recall status is terminated (terminated Jun 3, 2015). Affected quantity: 8.

Recall details

Product Type
Devices
Report Date
Aug 14, 2013
Initiation Date
Mar 14, 2013
Termination Date
Jun 3, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8

Product description

Rev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instruments which capture, hold and drive pedicle screw assemblies. There is a capturing feature that engages and holds the screw in place. There is also a feature used to lock the screw to the inserter before insertion of the screw. They can also be used to remove screws or back screws out. Screw inserters are used during the implantation of spinal pedicle screws intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. Used in conjunction with Pedicle Screw Spinal system marketed as KRD1 PedFuse Pedicle Screw System .

Reason for recall

Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the collar is not in place the locking sleeve can also slide off the inserter.

Distribution pattern

Distribution only in Texas.

Code information

Lot Number: PM3979

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1926-2013
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1926-2013. Report a data issue: hello@regdatalab.com.