Z-1926-2015 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-1926-2015 was initiated by Roche Diagnostics Operations, Inc. on May 7, 2015 and is designated Class II. Reason for recall: Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine… The recall status is terminated (terminated Mar 24, 2017). Affected quantity: 6255.

Recall Details

Product Type
Devices
Report Date
Jul 8, 2015
Initiation Date
May 7, 2015
Termination Date
Mar 24, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6255

Product Description

Lactate. Intended for use on COBAS INTEGRA 800 and 400 plus systems, cobas c111, Roche/Hitachi MODULAR systems, and cobas c systems for the quantitative determination of lactate concentration in plasm and cerebrospinal fluid.

Reason for Recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Distribution Pattern

US (nationwide) including PR.

Code Information

Catalog Numbers: 11822837190; 05401666190; 03183700190; 05171881190; 03183700190.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1926-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.