Z-1927-2026 Class II Ongoing

Recalled by Foundation Medicine, Inc. — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
April 29, 2026
Initiation Date
March 20, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
163 units

Product Description

Sequencing Agent SEQ0067, component of FoundationOne CDx (F1CDx)

Reason for Recall

Possible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variants

Distribution Pattern

US Nationwide distribution in the states of Massachusetts and North Carolina.

Code Information

Model Number: SEQ0067. Lot Numbers: 24346, 24386, 24488.