Z-1929-2025 Class II Ongoing
FDA device recall Z-1929-2025 was initiated by Philips North America on May 29, 2025 and is designated Class II. Reason for recall: identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a… The recall status is ongoing. Affected quantity: 1 unit.
Recall details
- Product Type
- Devices
- Report Date
- Jun 18, 2025
- Initiation Date
- May 29, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit
Product description
Intera 1.5T Achieva IT Nova Product Number: 781175
Reason for recall
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.
Distribution pattern
Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.
Code information
Product Number: 781175; UDI-DI: no UDI; Serial Number: 15003;
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1929-2025
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab