Z-1932-2020 Class II Terminated

Recalled by Philips North America, LLC — Andover, MA

FDA device recall Z-1932-2020 was initiated by Philips North America, LLC on Apr 10, 2020 and is designated Class II. Reason for recall: When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this… The recall status is terminated (terminated Jan 11, 2021). Affected quantity: 25 total.

Recall details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
Apr 10, 2020
Termination Date
Jan 11, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25 total

Product description

Philips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.

Reason for recall

When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.

Distribution pattern

Worldwide distribution - US Nationwide distribution including in the states of MD and MI. The countries of Austria, Belgium, France, Germany, Netherlands, Norway, South Africa, Spain, Switzerland, and United Kingdom.

Code information

Serial Numbers 17 7

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1932-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America, LLC →

Source: openFDA record for Z-1932-2020. Report a data issue: hello@regdatalab.com.