Z-1932-2020 Class II Terminated

Recalled by Philips North America, LLC — Andover, MA

Recall Details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
April 10, 2020
Termination Date
January 11, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25 total

Product Description

Philips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.

Reason for Recall

When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of MD and MI. The countries of Austria, Belgium, France, Germany, Netherlands, Norway, South Africa, Spain, Switzerland, and United Kingdom.

Code Information

Serial Numbers 17 7