Z-1940-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 20, 2020
- Initiation Date
- April 15, 2020
- Termination Date
- April 15, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10
Product Description
Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wire balloon catheter, which serves as the delivery system. The stent is positioned between two radiopaque marker bands, which are located inside the balloon at the proximal and distal tapers of the balloon.
Reason for Recall
The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased procedural time and additional intervention to remove a stent that dislodged within an unintended location.
Distribution Pattern
Domestic distribution to OH, IA, VA, and NC. Foreign distribution to Canada.
Code Information
Lot number 10153876