Z-1940-2020 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-1940-2020 was initiated by Cook Inc. on Apr 15, 2020 and is designated Class II. Reason for recall: The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased… The recall status is terminated (terminated Apr 15, 2024). Affected quantity: 10.

Recall details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
Apr 15, 2020
Termination Date
Apr 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10

Product description

Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wire balloon catheter, which serves as the delivery system. The stent is positioned between two radiopaque marker bands, which are located inside the balloon at the proximal and distal tapers of the balloon.

Reason for recall

The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased procedural time and additional intervention to remove a stent that dislodged within an unintended location.

Distribution pattern

Domestic distribution to OH, IA, VA, and NC. Foreign distribution to Canada.

Code information

Lot number 10153876

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1940-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cook Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Cook Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Cook Inc. →

Source: openFDA record for Z-1940-2020. Report a data issue: hello@regdatalab.com.