Z-1943-2024 Class II Ongoing

Recalled by Securitas Healthcare LLC — Lincoln, NE

Recall Details

Product Type
Devices
Report Date
June 12, 2024
Initiation Date
April 11, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
661 sites with mobile application user licenses

Product Description

Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5, a component of the Arial Emergency and Nurse Call System.

Reason for Recall

The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm notifications and other events when the app restarts, experience difficulty in placing and maintaining staff-to-staff phone calls, and frequently appears offline while the app restarts.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Canada, England, Hong Kong, France, Qatar, and Spain.

Code Information

Arial Mobile v11.0.0 UDI: (01)00840948100963(10)1100.