Z-1944-2012 Class II Terminated

Recalled by Impac Medical Systems Inc — Sunnyvale, CA

Recall Details

Product Type
Devices
Report Date
July 18, 2012
Initiation Date
June 8, 2012
Termination Date
June 19, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11 affected sites

Product Description

MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086

Reason for Recall

Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes after a machine terminates/faults, the remaining Wedge MU was being erroneously set to zero.

Distribution Pattern

Worldwide Distribution -- USA, including the states of VA, CT, GA, NJ and NY and the countries of Belgium, Denmark, France, and the United Kingdom.

Code Information

Version 2.10 - all builds and version 2.20 - builds up to and including 2.20.07A7.