FDA device recall Z-1949-2026 was initiated by Hologic, Inc on Mar 23, 2026 and is designated Class II. Reason for recall: It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications… The recall status is ongoing. Affected quantity: 1200 units.
Recall details
- Product Type
- Devices
- Report Date
- May 6, 2026
- Initiation Date
- Mar 23, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1200 units
Product description
Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.
Reason for recall
It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.
Code information
Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;