Z-1949-2026 Class II Ongoing

Recalled by Hologic, Inc — Marlborough, MA

Recall Details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
March 23, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1200 units

Product Description

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.

Reason for Recall

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.

Code Information

Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;