Z-1956-2026 Class II Ongoing

Recalled by Philips North America — Cambridge, MA

FDA device recall Z-1956-2026 was initiated by Philips North America on Apr 14, 2026 and is designated Class II. Reason for recall: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings… The recall status is ongoing. Affected quantity: 54 units.

Recall details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
Apr 14, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
54 units

Product description

Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115. 510(k) Numbers: K163116, K173079, K183063, K193215, K213583, K223458, K251397, K251808

Reason for recall

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Distribution pattern

Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.

Code information

1. Model Number (REF): 781315. UDI: (01)00884838099715(21)61067, (01)00884838099715(21)61058, (01)00884838099715(21)61056. Serial Numbers: 61067 61058, 61056. 2. Model Number (REF): 781341. UDI: (01)00884838055322(21)70229, (01)00884838055322(21)84667, (01)00884838055322(21)70311, (01)00884838055322(21)70668, (01)00884838055322(21)70845, (01)00884838055322(21)70851, (01)00884838055322(21)87151, (01)00884838055322(21)87112, (01)00884838055322(21)84322, (01)00884838055322(21)84323, (01)00884838055322(21)84228, (01)00884838055322(21)84789, (01)00884838055322(21)84261, (01)00884838055322(21)84500, (01)00884838055322(21)84556, (01)00884838055322(21)87187, (01)00884838055322(21)84840, (01)00884838055322(21)70685, (01)00884838055322(21)84661, (01)00884838055322(21)70787, (01)00884838055322(21)70063, (01)00884838055322(21)70621, (01)00884838055322(21)70743, (01)00884838055322(21)70911, (01)00884838055322(21)84049, (01)00884838055322(21)84088, (01)00884838055322(21)84147, (01)00884838055322(21)84254, (01)00884838055322(21)84283, (01)00884838055322(21)84403, (01)00884838055322(21)84433, (01)00884838055322(21)84460, (01)00884838055322(21)84462, (01)00884838055322(21)84467, (01)00884838055322(21)84475, (01)00884838055322(21)84734, (01)00884838055322(21)84538. Serial Numbers: 70229, 84667, 70311, 70668, 70845, 70851, 87151, 87112, 84322, 84323, 84228, 84789, 84261, 84500, 84556, 87187, 84840, 70685, 84661, 70787, 70063, 70621, 70743, 70911, 84049, 84088, 84147, 84254, 84283, 84403, 84433, 84460, 84462, 84467, 84475, 84498, 84538. 3. Model Number (REF): 781396. UDI: (01)00884838009820(21)41175, (01)00884838009820(21)41108, (01)00884838009820(21)41395, (01)00884838009820(21)41037, (01)00884838009820(21)41066, (01)00884838009820(21)41098, (01)00884838009820(21)41194, (01)00884838009820(21)41208, (01)00884838009820(21)41354, (01)00884838009820(21)41356, (01)00884838009820(21)41370, (01)00884838009820(21)41552. Serial Numbers: 41175, 41108, 41395, 41037, 41066, 41098, 41194, 41208, 41354, 41356, 41370, 41552. 4. Model Number (REF): 782115. UDI: (01)00884838099043(21)61214, (01)00884838099043(21)61221. Serial Numbers: 61214, 61221.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1956-2026