FDA device recall Z-1961-2014 was initiated by Integra LifeSciences Corp. on Jun 9, 2014 and is designated Class II. Reason for recall: Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially… The recall status is terminated (terminated Nov 6, 2014). Affected quantity: 94 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 9, 2014
- Initiation Date
- Jun 9, 2014
- Termination Date
- Nov 6, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 94 units
Product description
MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws
Reason for recall
Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.
Distribution pattern
Worldwide Distribution. US including states of CA, FL, IL, MD, MO, TN, TX, and countries of Australia, Czech Republic and Spain.
Code information
Lot No. Z21781A (45mm NewPort MIS Rod, 2) Z21956A (50mm NewPort MIS Rod, 2) Z21782A (55mm NewPort MIS Rod, 2) Z21957A (60mm NewPort MIS Rod, 2) Z21784A (75mm NewPort MIS Rod, 2) Z21958A (80mm NewPort MIS Rod, 2)