Z-1965-2016 Class II Terminated

Recalled by Integra LifeSciences Corp. — Plainsboro, NJ

FDA device recall Z-1965-2016 was initiated by Integra LifeSciences Corp. on May 10, 2016 and is designated Class II. Reason for recall: As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior… The recall status is terminated (terminated Dec 16, 2016). Affected quantity: 71 units.

Recall details

Product Type
Devices
Report Date
Jun 22, 2016
Initiation Date
May 10, 2016
Termination Date
Dec 16, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
71 units

Product description

TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.

Reason for recall

As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder

Distribution pattern

US Distribution to states of: CO, FL, IL, MI, NJ, NY, SC, TX, and VA.

Code information

Lot UP15H2, Lot UP09D3, Lot UP07E2

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1965-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1965-2016. Report a data issue: hello@regdatalab.com.