Z-1965-2016 Class II Terminated
FDA device recall Z-1965-2016 was initiated by Integra LifeSciences Corp. on May 10, 2016 and is designated Class II. Reason for recall: As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior… The recall status is terminated (terminated Dec 16, 2016). Affected quantity: 71 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 22, 2016
- Initiation Date
- May 10, 2016
- Termination Date
- Dec 16, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 71 units
Product description
TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.
Reason for recall
As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder
Distribution pattern
US Distribution to states of: CO, FL, IL, MI, NJ, NY, SC, TX, and VA.
Code information
Lot UP15H2, Lot UP09D3, Lot UP07E2