Z-1966-2012 Class II Terminated
FDA device recall Z-1966-2012 was initiated by Terumo Cardiovascular Systems Corporation on Jun 22, 2012 and is designated Class II. Reason for recall: Terumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery… The recall status is terminated (terminated Nov 8, 2013). Affected quantity: approx. 632 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 18, 2012
- Initiation Date
- Jun 22, 2012
- Termination Date
- Nov 8, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- approx. 632 units
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV. Internationally to: Republic, Egypt, Germany, Guatamala, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam,
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1966-2012
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude