Z-1970-2023 Class II Ongoing

Recalled by LumiraDx — Waltham, MA

FDA device recall Z-1970-2023 was initiated by LumiraDx on Apr 25, 2023 and is designated Class II. Reason for recall: Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention… The recall status is ongoing. Affected quantity: 1003 kits.

Recall details

Product Type
Devices
Report Date
Jun 21, 2023
Initiation Date
Apr 25, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1003 kits

Product description

SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB

Reason for recall

Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab

Distribution pattern

US Nationwide distribution.

Code information

GTIN-DI: 00850027193205 Pack Lot Number : 85438; Individual Swab Lot Number: 86445 Exp Date: 28-Oct-2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1970-2023
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1970-2023. Report a data issue: hello@regdatalab.com.