Z-1970-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 21, 2023
- Initiation Date
- April 25, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1003 kits
Product Description
SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB
Reason for Recall
Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab
Distribution Pattern
US Nationwide distribution.
Code Information
GTIN-DI: 00850027193205 Pack Lot Number : 85438; Individual Swab Lot Number: 86445 Exp Date: 28-Oct-2023