Z-1975-2026 Class I Ongoing
FDA device recall Z-1975-2026 was initiated by AVID Medical, Inc. on Mar 13, 2026 and is designated Class I. Reason for recall: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold. The recall status is ongoing. Affected quantity: 2,932 kits.
Recall details
- Product Type
- Devices
- Report Date
- May 13, 2026
- Initiation Date
- Mar 13, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,932 kits
Product description
Halyard HEART CATH, SELF REGIONAL kit. Model Number: SELF131-05.
Reason for recall
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Distribution pattern
US distribution to FL, IL, MO, MS, NC, NE, TX.
Code information
Model Number: SELF131-05. UDI-DI (Kit Lot Numbers): 10809160408650 (1611688, 1611689, 1618351, 1621441, 1627930, 1635908, 1644501, 1646021, 1651957, 1654887, 1656582, 1656581, 1659713, 1670884, 1670885, 1671950).
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1975-2026
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab