Z-1980-2025 Class II Ongoing
FDA device recall Z-1980-2025 was initiated by Ohio Medical Corporation on May 8, 2025 and is designated Class II. Reason for recall: Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect. The recall status is ongoing. Affected quantity: 36 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 25, 2025
- Initiation Date
- May 8, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 36 units
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.