Z-1982-2023 Class II Ongoing

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

FDA device recall Z-1982-2023 was initiated by MEDLINE INDUSTRIES, LP - Northfield on May 3, 2023 and is designated Class II. Reason for recall: Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO… The recall status is ongoing. Affected quantity: 851,328 units.

Recall details

Product Type
Devices
Report Date
Jun 28, 2023
Initiation Date
May 3, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
851,328 units

Product description

MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel

Reason for recall

Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of SINGAPORE, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES, DUBAI, CHINA, PANAMA, CANADA, SAUDI ARABIA, BERMUDA, COLOMBIA, MAURITIUS, KUWAIT.

Code information

a) REF MNK0005: GTIN 30884389161566, Lot Numbers: LMPTH0189, LMPTH0191, LMPTH0193, LMPTH0195, LMPTH0197, LMPTH0199, LMPTH0201, LMPTH0203, LMPTH0205, LMPTH0207, LMPTH0211, LMPTH0213; b) REF MNK0015: GTIN 40884389154558, Lot Numbers: LMPTH0188, LMPTH0190, LMPTH0192, LMPTH0196, LMPTH0198, LMPTH0200, LMPTH0202, LMPTH0204, LMPTH0206, LMPTH0210, LMPTH0210B, LMPTH0212

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1982-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for MEDLINE INDUSTRIES, LP - Northfield? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for MEDLINE INDUSTRIES, LP - Northfield: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about MEDLINE INDUSTRIES, LP - Northfield →

Source: openFDA record for Z-1982-2023. Report a data issue: hello@regdatalab.com.