Z-1989-2015 Class II Terminated

Recalled by Alliance Partners LLC — San Antonio, TX

Recall Details

Product Type
Devices
Report Date
July 15, 2015
Initiation Date
June 9, 2015
Termination Date
October 9, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21 units

Product Description

Nakoma-SL Anterior Cervical Plating System's Variable Single Barrel Drill Guide Product Usage: The Nakoma-SL ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The Nakoma ACP System is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusion in patients with the following indications: Degenerative Disk Disease, Trauma, Tumors, Deformities or curvatures, Psuedoarthrosis, Failed previous fusions, Spondylolistesis, and Spinal Stenosis.

Reason for Recall

Inadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw advancing through the plate during implantation.

Distribution Pattern

US Nationwide Distribution in the states LA, TX, and PR.

Code Information

Lot numbers SM59161 and SM60278