Z-1989-2020 Class II Terminated

Recalled by Covidien Llc — Mansfield, MA

Recall Details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
April 30, 2020
Termination Date
December 14, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8162 units

Product Description

Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P16TS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Reason for Recall

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Distribution Pattern

US Nationwide distribution.

Code Information

Lot Numbers: 1909308664, 1909320964 1933802964