Z-1990-2020 Class II Terminated

Recalled by Covidien Llc — Mansfield, MA

Recall Details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
April 30, 2020
Termination Date
December 14, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23587 units

Product Description

Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P16TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Reason for Recall

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Distribution Pattern

US Nationwide distribution.

Code Information

Lot Numbers: 1826714264, 1908533364, 1909302964, 1909303064, 1910654864, 1911300664, 1916240464, 1921106564, 1922533864, 2004908564, 2005617264, 2006339564