Z-1993-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 20, 2020
- Initiation Date
- April 30, 2020
- Termination Date
- December 14, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 120 units
Product Description
Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray,18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P18XTS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Reason for Recall
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Distribution Pattern
US Nationwide distribution.
Code Information
Lot Numbers: 1919708264