Z-1994-2013 Class II Terminated
FDA device recall Z-1994-2013 was initiated by Getinge USA Inc on Aug 2, 2013 and is designated Class II. Reason for recall: AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This could result in a false negative (or pass) when exposed to sub-lethal sterilization… The recall status is terminated (terminated Feb 23, 2017). Affected quantity: Catalog # 61301606636: 78 Packs (3900 BIs); Catalog # 61301606637: 43 Packs (4300 BIs).
Recall details
- Product Type
- Devices
- Report Date
- Aug 28, 2013
- Initiation Date
- Aug 2, 2013
- Termination Date
- Feb 23, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- Catalog # 61301606636: 78 Packs (3900 BIs); Catalog # 61301606637: 43 Packs (4300 BIs)
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA and the country of Canada.