Z-1994-2020 Class II Terminated

Recalled by Covidien Llc — Mansfield, MA

Recall Details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
April 30, 2020
Termination Date
December 14, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
0

Product Description

Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray, 18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P18XTSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Reason for Recall

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Distribution Pattern

US Nationwide distribution.

Code Information

Lot Numbers: 1914133564