Z-1999-2016 Class II Terminated

Recalled by Siemens Medical Solutions USA Inc. — Knoxville, TN

FDA device recall Z-1999-2016 was initiated by Siemens Medical Solutions USA Inc. on Apr 25, 2016 and is designated Class II. Reason for recall: The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service. The recall status is terminated (terminated Feb 1, 2017). Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Jun 22, 2016
Initiation Date
Apr 25, 2016
Termination Date
Feb 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10528956 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Reason for recall

The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of China and Malaysia.

Code information

Serial Number 11009

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1999-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA Inc. →

Source: openFDA record for Z-1999-2016. Report a data issue: hello@regdatalab.com.