Z-2002-2016 Class II Terminated

Recalled by Datascope Corporation — Fairfield, NJ

FDA device recall Z-2002-2016 was initiated by Datascope Corporation on May 9, 2016 and is designated Class II. Reason for recall: Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits. The recall status is terminated (terminated Mar 28, 2017). Affected quantity: 17 units.

Recall details

Product Type
Devices
Report Date
Jun 22, 2016
Initiation Date
May 9, 2016
Termination Date
Mar 28, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17 units

Product description

Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the hear-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours

Reason for recall

Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.

Distribution pattern

US in the state of GA

Code information

Lot/batch No: 3000003260, Part number 701050175

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2002-2016
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2002-2016. Report a data issue: hello@regdatalab.com.