Z-2002-2016 Class II Terminated
FDA device recall Z-2002-2016 was initiated by Datascope Corporation on May 9, 2016 and is designated Class II. Reason for recall: Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits. The recall status is terminated (terminated Mar 28, 2017). Affected quantity: 17 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 22, 2016
- Initiation Date
- May 9, 2016
- Termination Date
- Mar 28, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 17 units
Product description
Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the hear-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours
Reason for recall
Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.
Distribution pattern
US in the state of GA
Code information
Lot/batch No: 3000003260, Part number 701050175