Z-2003-2020 Class II Terminated

Recalled by Inpeco S.A. — Lugano, N/A

FDA device recall Z-2003-2020 was initiated by Inpeco S.A. on Apr 7, 2020 and is designated Class II. Reason for recall: When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples. The recall status is terminated (terminated Jul 28, 2022). Affected quantity: 82.

Recall details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
Apr 7, 2020
Termination Date
Jul 28, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
82

Product description

Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Reason for recall

When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.

Distribution pattern

Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansas, Hawaii, Arizona, Washington

Code information

Serial Numbers: AML.001, FLX.0104, FLX.0105, FLX.0106, FLX.0107, FLX.0108, FLX.0112, FLX.0114, FLX.0115, FLX.0116, FLX.0118, FLX.0119, FLX.0120, FLX.0124, FLX.0125, FLX.0126, FLX.0131, FLX.0133, FLX.0134, FLX.0135, FLX.0136, FLX.0140, FLX.0141, FLX.0142, FLX.0143, FLX.015, FLX.0152, FLX.0155, FLX.0158, FLX.0161, FLX.0163, FLX.0164, FLX.017, FLX.0175, FLX.0178, FLX.018, FLX.0184, FLX.0189, FLX.019, FLX.0196, FLX.020, FLX.021, FLX.023, FLX.026, FLX.027, FLX.030, FLX.032, FLX.033, FLX.038, FLX.040, FLX.044, FLX.045, FLX.047, FLX.051, FLX.052, FLX.053, FLX.054, FLX.055, FLX.058, FLX.059, FLX.063, FLX.064, FLX.065, FLX.066, FLX.067, FLX.068, FLX.071, FLX.072, FLX.074, FLX.075, FLX.076, FLX.077, FLX.079, FLX.080, FLX.081, FLX.091, FLX.092, FLX.095, FLX.099, FLX.101, FLX.102, FLX.103

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2003-2020
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Inpeco S.A.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2003-2020. Report a data issue: hello@regdatalab.com.