Z-2004-2020 Class II Terminated
FDA device recall Z-2004-2020 was initiated by Inpeco S.A. on Apr 7, 2020 and is designated Class II. Reason for recall: When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples. The recall status is terminated (terminated Jul 28, 2022). Affected quantity: 273.
Recall details
- Product Type
- Devices
- Report Date
- May 20, 2020
- Initiation Date
- Apr 7, 2020
- Termination Date
- Jul 28, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 273
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansas, Hawaii, Arizona, Washington
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2004-2020
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
Need the full FDA record for Inpeco S.A.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.