Z-2008-2023 Class II Ongoing

Recalled by Covidien, LLC — Minneapolis, MN

FDA device recall Z-2008-2023 was initiated by Covidien, LLC on May 22, 2023 and is designated Class II. Reason for recall: Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service… The recall status is ongoing. Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Jul 5, 2023
Initiation Date
May 22, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

Reason for recall

Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.

Distribution pattern

International distribution in the country of Taiwan.

Code information

UDI-DI: 10884521782211; Serial Number: 518503

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2008-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Covidien, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2008-2023. Report a data issue: hello@regdatalab.com.