Z-2008-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
March 19, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4225160 units

Product Description

Medline Neuro Sponges, various dimensions, labeled as: 1. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.25" X 0.25", XR, 10/PK, Medline SKU NEUROSPNG02; 2. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 0.5", XR, 10/PK, Medline SKU NEUROSPNG05; 3. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG06; 4. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1.5", XR, LF, 10/PK, Medline SKU NEUROSPNG07; 5. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 3", XR, LF, 10/PK, Medline SKU NEUROSPNG09; 6. Neuro Sponges, Eponges neuro, Neuroesponjas, size .75" X .75", XR, LF, 10/PK, Medline SKU NEUROSPNG11; 7. Neuro Sponges, Eponges neuro, Neuroesponjas, size 1" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG13; 8. Neuro Sponges, Eponges neuro, Neuroesponjas, size 1" X 3", XR, LF, 10/PK, Medline SKU NEUROSPNG14.

Reason for Recall

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Distribution Pattern

Worldwide distribution.

Code Information

1. NEUROSPNG02: UDI/DI each 10197344111572, UDI/DI case 20197344111579, ALL LOTS; 2. NEUROSPNG05: UDI/DI each 20080196286659, UDI/DI case 30080196286656, ALL LOTS; 3. NEUROSPNG06: UDI/DI each 20080196286505, UDI/DI case 30080196286502, ALL LOTS; 4. NEUROSPNG07: UDI/DI each 10080196286485, UDI/DI case 40080196286486, ALL LOTS; 5. NEUROSPNG09: UDI/DI each 20080196286536, UDI/DI case 30080196286533, ALL LOTS; 6. NEUROSPNG11: UDI/DI each 20080196286550, UDI/DI case 30080196286557, ALL LOTS; 7. NEUROSPNG13: UDI/DI each 10080196286577, UDI/DI case 40080196286578, ALL LOTS; 8. NEUROSPNG14: UDI/DI each 10080196286584, UDI/DI case 40080196286585, ALL LOTS.