Z-2009-2012 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-2009-2012 was initiated by Terumo Cardiovascular Systems Corporation on Jun 21, 2012 and is designated Class II. Reason for recall: Terumo Cardiovascular Systems has received reports of several issues that manifest themselves as communications or power failures in the system network for Terumo Advanced Perfusion System 1. Issues… The recall status is terminated (terminated Jun 10, 2013). Affected quantity: 992 units.

Recall details

Product Type
Devices
Report Date
Jul 25, 2012
Initiation Date
Jun 21, 2012
Termination Date
Jun 10, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
992 units

Product description

Terumo Advanced Perfusion System 1 System 1 Base, 100/120V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Reason for recall

Terumo Cardiovascular Systems has received reports of several issues that manifest themselves as communications or power failures in the system network for Terumo Advanced Perfusion System 1. Issues that affect a single pump or module are: " The appearance of a red X on a single pump or module icon on the CCM " The appearance or a single '?' symbol on a pump or module icon on the CCM " The fai

Distribution pattern

Worldwide Distribution-USA (nationwide) including the states of , AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV and the countries of ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.

Code information

Catalog number: 801763 and serial number: 0011-0335,0337-0610, 1001-1034, 1037-1049,1100, 1103-1106, 1109-1445,and 1447-1450.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2009-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-2009-2012. Report a data issue: hello@regdatalab.com.