Z-2013-2025 Class I Ongoing

Recalled by Maquet Critical Care AB — Solna

FDA device recall Z-2013-2025 was initiated by Maquet Critical Care AB on May 30, 2025 and is designated Class I. Reason for recall: Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier. The recall status is ongoing. Affected quantity: 874 units.

Recall details

Product Type
Devices
Report Date
Jul 9, 2025
Initiation Date
May 30, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
874 units

Product description

Servo-u Ventilator System. Model Number: 6688600.

Reason for recall

Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.

Distribution pattern

US Nationwide distribution and Puerto Rico. No international distribution.

Code information

Model Number: 6688600. UDI-DI: 07325710001042. All software versions.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2013-2025
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Maquet Critical Care AB? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2013-2025. Report a data issue: hello@regdatalab.com.