Z-2025-2019 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
July 31, 2019
Initiation Date
June 12, 2019
Termination Date
September 18, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19

Product Description

AXIOM Artis dFC, Model Number 7727717

Reason for Recall

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Distribution Pattern

US Nationwide Distribution

Code Information