Z-2027-2019 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
July 31, 2019
Initiation Date
June 12, 2019
Termination Date
September 18, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
53

Product Description

AXIOM Artis dBC, Model Number 7728392

Reason for Recall

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Distribution Pattern

US Nationwide Distribution

Code Information

Serial Numbers: 44125 44151 44101 44246 44185 44262 44263 44199 44105 44084 44080 44072 44243 44198 44171 44275 44215 44255 44081 44242 44265 44189 44289 44145 44060 44144 44184 44286 44223 44068 44193 44244 44201 44251 44229 44304 44152 44136 44254 44277 44279 44268 44206 44278 44259 44260 44168 44267 44110 44218 44165 44271 44135