Z-2027-2023 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 12, 2023
- Initiation Date
- May 19, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 36337 devices
Product Description
TruSignal AllFit Sensor: a) REF TS-AF-10, 10 pcs; B) REF AF-25, 25 pcs; Oximeter
Reason for Recall
There is a potential reduction of energy reaching patient during defibrillation, potential contact with unintended voltage, or inaccurate measurement.
Distribution Pattern
Worldwide
Code Information
a) REF TS-AF-10, GTIN 0840682103176 b) REF TS-AF-25, GTIN 0840682103176