Z-2042-2026 Class II Ongoing
FDA device recall Z-2042-2026 was initiated by Medline Industries, LP on March 16, 2026 and is designated Class II. Reason for recall: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Pr… The recall status is ongoing. Affected quantity: 1590 kits.
Recall Details
- Product Type
- Devices
- Report Date
- May 6, 2026
- Initiation Date
- March 16, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1590 kits
Product Description
Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: LATEX FREE GENERAL PURPOSE TRAY, Medline Kit SKU 75670
Reason for Recall
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Distribution Pattern
US Nationwide distribution.
Code Information
UDI/DI each 00653160095890, UDI/DI case 10653160095897, Lot Number: 2025102290