Z-2042-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-2042-2026 was initiated by Medline Industries, LP on Mar 16, 2026 and is designated Class II. Reason for recall: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol… The recall status is ongoing. Affected quantity: 1590 kits.

Recall details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
Mar 16, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1590 kits

Product description

Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: LATEX FREE GENERAL PURPOSE TRAY, Medline Kit SKU 75670

Reason for recall

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

Distribution pattern

US Nationwide distribution.

Code information

UDI/DI each 00653160095890, UDI/DI case 10653160095897, Lot Number: 2025102290

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2042-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2042-2026. Report a data issue: hello@regdatalab.com.