Z-2043-2015 Class II Terminated
FDA device recall Z-2043-2015 was initiated by Volcano Corp on Sep 23, 2013 and is designated Class II. Reason for recall: A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction… The recall status is terminated (terminated Jul 9, 2015). Affected quantity: 227 total units affected by the issue - all models.
Recall details
- Product Type
- Devices
- Report Date
- Jul 15, 2015
- Initiation Date
- Sep 23, 2013
- Termination Date
- Jul 9, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 227 total units affected by the issue - all models
Product description
Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE MOBILE - 400-0100.01. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.
Reason for recall
A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.
Distribution pattern
Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.
Code information
s5 Part number 804200-002 and 804200-014; s5x Part number 807300-001; CORE MOBILE Part Number 400-0100.01. All software versions