Z-2066-2026 Class I Ongoing

Recalled by Windstone Medical Packaging, Inc. — Billings, MT

FDA device recall Z-2066-2026 was initiated by Windstone Medical Packaging, Inc. on Apr 2, 2026 and is designated Class I. Reason for recall: Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full… The recall status is ongoing. Affected quantity: 150 kits (US only).

Recall details

Product Type
Devices
Report Date
May 20, 2026
Initiation Date
Apr 2, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
150 kits (US only)

Product description

Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F

Reason for recall

Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.

Distribution pattern

US Nationwide distribution in the state of Arizona.

Code information

Pack AMS6908E UDI-DI B098AMS6908E0 Lot 214875 Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2066-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2066-2026. Report a data issue: hello@regdatalab.com.