Z-2067-2019 Class I Terminated

Recalled by Medline Industries Inc — Northfield, IL

Recall Details

Product Type
Devices
Report Date
August 21, 2019
Initiation Date
May 10, 2019
Termination Date
August 18, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4890

Product Description

Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.

Reason for Recall

This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.

Distribution Pattern

One distributor located in Illinois.

Code Information

lots 2018090590 2018112190 2018090690 2018101190 2018101590 2018121190 2019013190 2019030490 2019040190