Z-2068-2017 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
May 17, 2017
Initiation Date
March 22, 2017
Termination Date
October 3, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8154

Product Description

Regenerex Patella RGX 3 PEG SER A PATELLA 28MM RGX 3 PEG SER A PATELLA 31MM RGX 3 PEG SER A PATELLA 34MM RGX 3 PEG SER A PATELLA 37MM Product Usage: The Regenerex Series A Patella can be used for any non-cemented resurfaced 3-peg patella application within the Vanguard Complete Knee System. 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved. 2. Correction of varus, valgus, or posttraumatic deformity. 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

Reason for Recall

pegs shearing post-operatively

Distribution Pattern

US Nationwide.distribution and Canada.

Code Information

All lots of Model #'s: 141355 141356 141357 141358