Z-2073-2019 Class I Terminated

Recalled by St Jude Medical Inc. — Sylmar, CA

FDA device recall Z-2073-2019 was initiated by St Jude Medical Inc. on Jun 20, 2019 and is designated Class I. Reason for recall: Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The… The recall status is terminated (terminated Apr 3, 2024). Affected quantity: 21.

Recall details

Product Type
Devices
Report Date
Aug 14, 2019
Initiation Date
Jun 20, 2019
Termination Date
Apr 3, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21

Product description

ELLIPSE DR, REF: CD2377-36QC Implantable cardioverter defibrillators (ICDs)

Reason for recall

Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

Distribution pattern

Worldwide distribution. US nationwide, and countries: DE, FR, IT, ES, NL, PT, SE, HU, DK, PL, IL, GP, DZ, CH, SA, SK, LU, NZ, VN, and NO

Code information

Serial Number: 1282011, 1282019, 1282065, 1282069, 1282424, 1282458, 1282469, 1282474, 1282596, 1282602, 1282609, 1282622, 1282625, 1282638, 1282641, 1282648, 9829389, 9831310, 9831509, 9831515, 9831516

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2073-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for St Jude Medical Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for St Jude Medical Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about St Jude Medical Inc. →

Source: openFDA record for Z-2073-2019. Report a data issue: hello@regdatalab.com.