Z-2076-2019 Class II Terminated

Recalled by LABORIE MEDICAL TECHNOLOGIES, CORP — Williston, VT

FDA device recall Z-2076-2019 was initiated by LABORIE MEDICAL TECHNOLOGIES, CORP on Jul 17, 2019 and is designated Class II. Reason for recall: Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire… The recall status is terminated (terminated Jul 10, 2023). Affected quantity: 2 US.

Recall details

Product Type
Devices
Report Date
Jul 31, 2019
Initiation Date
Jul 17, 2019
Termination Date
Jul 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 US

Product description

U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions

Reason for recall

Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard

Distribution pattern

US distribution in AZ, and PA.

Code information

Serial Numbers: PE18-7SBL1452, PE18-7SBL 1414